GLP hERG testing assay validation following ICH E14/S7B 2022 Q&A best practice guidelines

Authors

James Masterson, Sergiy Tokar, Ayesha Jinat, Robert Kirby

Metrion Biosciences Ltd, Granta Centre, Granta Park, Cambridge, CB21 6AL, United Kingdom

Introduction

Importance of GLP hERG testing for potassium ion channels

Good Laboratory Practice (GLP) hERG testing is a critical component in the preclinical safety evaluation of new chemical entities. The human ether-à-go-go-related gene (hERG) encodes a key potassium ion channel that plays an essential role in the repolarisation phase of the cardiac action potential. Any alterations in the function of these channels can lead to severe cardiac arrhythmias and potentially fatal outcomes. As such, a robust hERG assay is indispensable in predicting cardiotoxicity risk. This publication details the validation of Metrion’s GLP hERG testing assay in strict accordance with the ICH E14/S7B 2022 Q&A best practice guidelines, ensuring that our methods deliver high-quality, reproducible, and transparent data.

Overview

ICH E14/S7B 2022 Q&A best practice guidelines

The ICH E14/S7B 2022 Q&A guidelines provide comprehensive recommendations on evaluating the effects of preclinical compounds on hERG potassium ion channels. These guidelines ensure that the in vitro hERG assessments are carried out in full GLP compliance, thereby bolstering data integrity and regulatory acceptance. The guidelines recommend specific experimental methods, quality control measures, and reporting formats designed to achieve industry-wide consistency and transparency.

A central tenet of the guidelines is the use of positive controls - ondansetron, moxifloxacin, and dofetilide - to establish defined safety margins. In our study, a hERG assay is deemed negative if the test article’s safety margin exceeds those derived from these controls. This framework is essential to both industry best practice and regulatory compliance.

Materials and Methods

Detailed materials and methods for the hERG assay

Cell Preparation: Culturing Potassium Ion Channel-Expressing Cells

The cornerstone of our GLP hERG testing protocol is the use of Chinese hamster ovary (CHO) cells stably expressing the hERG1a isoform. These cells are well recognised for their consistent and robust expression of potassium ion channels, making them ideal for electrophysiological studies. The cells were carefully plated onto glass coverslips and maintained at 37 °C for 24 to 48 hours prior to electrophysiological recordings. This incubation period ensures optimal cell adhesion and viability, which is critical for obtaining reliable hERG assay data.

Optimised recording solutions

To faithfully mimic the physiological environment during recordings, specialised intracellular and extracellular solutions were prepared:

  • Intracellular Solution (in mM): 120 K-gluconate, 20 KCl, 10 HEPES, 5 EGTA, 5 MgATP; adjusted to a pH of 7.3. This solution is formulated to replicate the intracellular ionic milieu, ensuring stability during electrophysiological measurements.
  • Extracellular Solution (in mM): 130 NaCl, 10 HEPES, 5 KCl, 1 MgCl₂·6H₂O, 1 CaCl₂·2H₂O, 12.5 dextrose; adjusted to a pH of 7.4. The extracellular solution provides the necessary conditions for normal channel activity and maintains cell viability throughout the experiment.

Compound preparation for hERG channel testing

For a robust hERG assay, positive control compounds must be precisely prepared:

  • Moxifloxacin: Prepared by dissolving in the extracellular solution. Stock solutions were stored at -20 °C, with a fresh aliquot thawed each day to ensure accuracy in concentration.
  • Ondansetron and Dofetilide: These compounds were dissolved in 100% dimethyl sulfoxide (DMSO) and stored at -20 °C. A fresh aliquot was thawed daily and diluted in the extracellular solution to achieve a final DMSO concentration of 0.1%, thereby maintaining consistency across tests.

Electrophysiological recording using the manual patch clamp technique

The manual patch clamp technique is considered the gold standard for GLP hERG testing. Our whole-cell voltage-clamp experiments were conducted using a HEKA EPC10 patch clamp amplifier paired with PatchMaster Pro (version V2.73) software. Recordings were carried out at 36 ± 1 °C to closely replicate human physiological conditions. Key details include:

  • Minimising Voltage Errors: Voltage errors were kept to a minimum by utilising ≥ 80% series resistance (Rs) compensation.
  • Sampling and Filtering: Membrane currents were sampled at 20 kHz and then filtered using a dual-stage low pass Bessel filter set at 10 kHz and 2.9 kHz, respectively. This approach ensures that the electrophysiological signals are captured accurately, preserving the integrity of the hERG assay data.
  • Voltage Protocol Implementation: A voltage protocol, based on the “Recommended voltage protocol to study drug-cardiac ion channel interactions using recombinant cell lines” (FDA 2021 guidelines), was applied at 0.2 Hz. The voltage command was adjusted for the liquid junction potential (-15 mV) to ensure precise current measurements.
The voltage protocol that was used to elicit hERG current

Figure 1. The voltage protocol that was used to elicit hERG current. Shaded areas indicate the locations of the cursors used to analyse the holding current (green), the current elicited by a step to -90 mV (red), and the peak current (blue).

Compound application and quality control

Following the establishment of the whole-cell configuration, cells were initially superfused with extracellular solution containing the vehicle (0.1% DMSO) to allow the current to stabilise during the control period. After stabilisation, compounds were applied using either a two-point assay format (for ondansetron and moxifloxacin) or a single-point assay format (for dofetilide). Each experiment was concluded with the application of 1 µM E-4031, which permitted the subtraction of leak and endogenous currents from the recorded data.

Quality control was rigorously maintained by continuously monitoring the peak current amplitude, input resistance, series resistance, and holding current over time. Cells exhibiting significant fluctuations were discarded to ensure that only high-quality data were used in the analysis. Additionally, samples of the compounds were collected after each experiment to verify concentration accuracy, further enhancing the reliability of our hERG assay.

The compound application paradigms
Figure 2. Compound application paradigms

Figure 2. Paradigms used for compound application, including two-point assay formats for ondansetron and moxifloxacin (Figure 2A) and a single-point assay format for dofetilide (Figure 2B), as well as vehicle controls (Figure 2C).

Data analysis: Quantifying hERG channel inhibition

Data analysis in GLP hERG testing is critical for determining the safety margins of test compounds. In our study, the percentage inhibition of the peak current amplitude was calculated using values derived after subtracting the effects of 1 µM E-4031. The mean inhibition data (± standard deviation) were plotted against the logarithm of the compound concentrations, and the data were fitted using a four-parameter logistic equation. This analysis allowed us to derive in-house IC50 values for each compound, complete with 95% confidence intervals.

This robust data analysis protocol ensures that the hERG assay is not only reproducible but also adheres strictly to the quality control measures dictated by the ICH E14/S7B guidelines. The successful validation of the assay using ondansetron, moxifloxacin, and dofetilide confirms that our GLP hERG testing protocol reliably measures the impact on potassium ion channels, offering an indispensable tool in the preclinical safety assessment of investigational drugs.

  • Percentage inhibition of the peak current amplitude was calculated using E-4031 subtracted current values.
  • The mean (± SD) inhibition data were plotted against concentration (on a logarithmic scale) and fitted with the following four-parameter logistic fit equation (A = 0, B = 100, C = pIC50, D = Hill slope):
GLP hERG data analysis equation
  • In addition, 95% confidence bands were included on each graph, as well as the individual inhibition values calculated for each concentration of the compounds.

Results

Demonstrating reliable in-house IC50 values

Representative electrophysiological data

Our GLP hERG testing assay produced high-resolution current traces that clearly demonstrate the inhibitory effects of the positive control compounds on hERG channels. These recordings allowed for the precise quantification of the effects on peak current amplitude, input resistance, series resistance, and holding current.

Representative current traces obtained from experiments with ondansetron, moxifloxacin, and dofetilide, alongside vehicle controls are shown in Figure 3. This figure also includes key electrophysiological parameters such as peak current amplitude and input resistance, reinforcing the consistency of our data.

Representative current traces

Figure 3. Representative current traces obtained from experiments with ondansetron, moxifloxacin, and dofetilide, alongside vehicle controls.

To further support our findings, we include two tables that summarise critical electrophysiological parameters and IC50 values derived from our GLP hERG testing assay.

Table 1: The mean (± SEM) Peak Current amplitude, Input Resistance, Series Resistance and Holding Current values recorded at the end of each Control Period.

Ondansetron Moxifloxacin Dofetilide Vehicle
Peak Current (pA) 1535.3 ± 242.5 1257.9 ± 119.1 1360.2 ± 170.9 1324.1 ± 242.4
Input Resistance (MΩ) 770 ± 90 580 ± 30 560 ± 50 680 ± 80
Series Resistance (MΩ) 3.5 ± 2.1 4.0 ± 3.0 3.2 ± 2.1 3.2 ± 1.3
Holding Current (pA) -32.1 ± 5.5 -37.1 ± 6.4 -25.5 ± 3.6 -18.4 ± 6.9

Table 2 compares the IC50 values and 95% confidence intervals reported in the ICH E14/S7B training material with those obtained by Metrion in this study for moxifloxacin, ondansetron, and dofetilide.

Compound ICH E14/S7B Training Material IC50 (µM) (95% Confidence Intervals) Metrion IC50 (µM) (95% Confidence Intervals)
Moxifloxacin 62 (38, 104) 96.2 (78.6, 117.7)
Ondansetron 1.4 (0.8, 2.6) 1.72 (1.51, 1.95)
Dofetilide 0.01 (<0.01, 0.02) 0.012 (0.011, 0.013)

Consistency with ICH E14/S7B training material

The IC50 values obtained in our study were within a twofold range of those reported in the ICH E14/S7B training material. This consistency confirms that our hERG assay, conducted in full GLP compliance, provides a reliable benchmark for evaluating test articles. The data underscore the importance of using well-established positive controls - ondansetron, moxifloxacin, and dofetilide - in determining the safety margins for potential cardiotoxic compounds.

Summary

Advancing drug safety through rigorous hERG assays

Enhancing preclinical assessment of potassium ion channels

The validation of our GLP hERG testing assay has significant implications for the pharmaceutical industry. By rigorously adhering to the ICH E14/S7B Q&A best practice guidelines, we have established a robust framework for evaluating the effects of preclinical compounds on potassium ion channels. The detailed electrophysiological data generated by our manual patch clamp technique provide clear insights into the safety profiles of investigational drugs, thereby reducing the risk of drug-induced cardiac arrhythmias.

Importance of GLP compliance in drug development

GLP compliance is not only a regulatory requirement but also a hallmark of quality in preclinical testing. By ensuring that our hERG assay is conducted under GLP conditions, we can guarantee the transparency, consistency, and reliability of our data. This commitment to GLP hERG testing reinforces our position as a leader in the field of ion channel research and supports safer drug development practices.

Future perspectives for hERG assays and potassium ion channel research

Looking ahead, there is considerable potential for integrating emerging automated patch clamp systems with traditional manual techniques. Such innovations promise to increase throughput while preserving the high fidelity of electrophysiological recordings. Continued refinement of GLP hERG testing protocols will further enhance our understanding of potassium ion channels, ultimately contributing to the development of safer therapeutic agents.

Conclusions

Setting a benchmark in GLP hERG testing and ion channel research

In summary, our GLP hERG testing assay validation - conducted in line with the ICH E14/S7B 2022 Q&A best practice guidelines - represents a significant advancement in the preclinical assessment of drug safety. Through meticulous cell preparation, optimised recording solutions, precise compound handling, and rigorous data analysis, we have demonstrated that our hERG assay provides reliable and reproducible measurements of potassium ion channel activity.

The successful validation of the assay using positive controls such as ondansetron, moxifloxacin, and dofetilide confirms its robustness and accuracy. This work not only enhances the overall reliability of preclinical testing but also sets a new benchmark for GLP hERG testing within the industry.

For organisations committed to the highest standards of drug safety, our publication provides a comprehensive guide on implementing best practice methodologies. By prioritising quality control and regulatory compliance, we continue to advance the field of potassium ion channel research, ultimately contributing to the development of safer and more effective therapeutic agents.

This detailed report is an essential resource for scientists and regulatory professionals alike, offering valuable insights into the optimisation of hERG assays and GLP hERG testing. Through our unwavering commitment to excellence, Metrion continues to lead the way in preclinical electrophysiological research, ensuring that innovative drugs are developed with the highest safety standards in mind.

References

Recommended Publications
Latest Publications
Development of a High-Throughput Automated Electrophysiology Assay for Human NaV1.9 Inhibitor Screening in Neuropathic Pain Research

Robust high-throughput automated electrophysiology assay using a monoclonal CHO-hNav1.9 cellular reagent suitable for fully supporting a Nav1.9 discovery program.

Development and Evaluation of Novel Solution Pairs to Enhance Seal Resistance in Automated Patch Clamp Electrophysiology and Ion Channel Experiments

To overcome seal enhancer limitations, Sophion and Metrion collaborated to determine whether other insoluble salts can act as seal enhancers and how these solution pairs affect the biophysical properties and pharmacology of the investigated ion channels.

View All
Metrion Biosciences is a contract research organisation (CRO) specialising in high-quality preclinical drug discovery services.
magnifier
linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram